Copyright © 1989 by British Editorial Society of Bone and Joint Surgery
Anterior cruciate ligament replacement. Knee stability and the effects of implants
Mechanical Engineering Department, Imperial College of Science, Technology and Medicine, London, England.
The anteroposterior stability of cadaveric knees was investigated. There was a wide range of normal laxity; knees were more stable at 90 degrees than at 20 degrees flexion. Anterior cruciate ligament implants with different stiffnesses were inserted; normal stability could always be restored, and the stiffness or extensibility of implants did not affect knee behaviour significantly. The tightness of implants was critical--small tensioning errors caused subluxation, inhibited knee extension and allowed damagingly high implant tensions. It is concluded that the tension of ligament implants could not be adjusted simply with a pre-set instrument; the procedure will remain critically dependent on the judgment of the operating surgeon. |
|