Logo of The Journal of Bone & Joint Surgery (Br)
Joint Replacement Instrumentation Limited (JRI) Ad
Quick search:        
          Advanced Search
Guest Access | Sign In
This Article
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow My Folders
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Engh, G.
Right arrow Articles by Hanes, C.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Engh, G.
Right arrow Articles by Hanes, C.
Journal of Bone and Joint Surgery - British Volume, Vol 74-B, Issue 1, 9-17
Copyright © 1992 by British Editorial Society of Bone and Joint Surgery


Articles

Polyethylene wear of metal-backed tibial components in total and unicompartmental knee prostheses

GA Engh, KA Dwyer, and CK Hanes

Anderson Orthopaedic Research Institute, Arlington, Virginia 22206.

We examined 86 polyethylene inserts, retrieved from total and unicompartmental knee prostheses after an average of 39.5 months in situ, grading them from 0 to 3 for seven modes of polyethylene degradation. Severe wear, with delamination or deformation, was observed in 51% of the implants, and was associated with time in situ, lack of congruency, thin polyethylene, third-body wear debris, and heat-pressed polyethylene. Significant under-surface cold flow was identified in some areas of unsupported polyethylene, and was associated with delamination in the load-bearing areas of thin inserts above screw holes in the underlying metal tray. We recommend the use of thicker polyethylene inserts, particularly in young, active patients and in designs with screw holes in the tibial baseplate. Thin polyethylene inserts which are at risk for accelerated wear and premature failure should be monitored radiographically at annual intervals.


This article has been cited by other articles:


Home page
J Bone Joint Surg BrHome page
A. J. Price, A. Short, C. Kellett, D. Beard, H. Gill, H. Pandit, C. A. F. Dodd, and D. W. Murray
Ten-year in vivo wear measurement of a fully congruent mobile bearing unicompartmental knee arthroplasty
J Bone Joint Surg Br, November 1, 2005; 87-B(11): 1493 - 1497.
[Abstract] [Full Text] [PDF]


Home page
J Bone Joint Surg BrHome page
J. Arora and A. C. Ogden
Osteolysis in a surface-cemented, primary, modular Freeman-Samuelson total knee replacement
J Bone Joint Surg Br, November 1, 2005; 87-B(11): 1502 - 1506.
[Abstract] [Full Text] [PDF]



(c) British Editorial Society of Bone and Joint Surgery All Rights Reserved
Registered charity no: 209299     Print ISSN: 0301-620X
Hip, Knee, Trauma, Upper limb, Foot & Ankle, Paediatrics, Oncology, Spine, Arthroplasty, General