Logo of The Journal of Bone & Joint Surgery (Br)
Joint Replacement Instrumentation Limited (JRI) Ad
Quick search:        
          Advanced Search
Guest Access | Sign In
This Article
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow My Folders
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Curtis, M.
Right arrow Articles by Hungerford, D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Curtis, M.
Right arrow Articles by Hungerford, D.
Journal of Bone and Joint Surgery - British Volume, Vol 74-B, Issue 3, 372-376
Copyright © 1992 by British Editorial Society of Bone and Joint Surgery


Articles

The initial stability of uncemented acetabular components

MJ Curtis, RH Jinnah, VD Wilson, and DS Hungerford

Johns Hopkins University, Baltimore, Maryland.

The osseo-integration of an uncemented acetabular component depends on its initial stability. This is usually provided by under-reaming of the acetabulum. We have assessed the fixation of 52 mm porous-coated hemispherical prostheses inserted into cadaveric acetabula under-reamed by 1, 2, 3 and 4 mm. We tested the torsional stability of fixation, after preloading with 686 N in compression, by measuring the torque required to produce 1 degree and 2 degrees of rotation. Under-reaming by 2 mm and 3 mm gave significantly better fixation than 1 mm (p less than 0.01, p less than 0.02). Insertion after under-reaming of 4 mm caused some fractures. To obtain maximum interference fit and optimal implant stability, we recommend the use of an implant 2 mm or 3 mm larger than the last reamer.


This article has been cited by other articles:


Home page
J Bone Joint Surg BrHome page
I. Udofia, F. Liu, Z. Jin, P. Roberts, and P. Grigoris
The initial stability and contact mechanics of a press-fit resurfacing arthroplasty of the hip
J Bone Joint Surg Br, April 1, 2007; 89-B(4): 549 - 556.
[Abstract] [Full Text] [PDF]



(c) British Editorial Society of Bone and Joint Surgery All Rights Reserved
Registered charity no: 209299     Print ISSN: 0301-620X
Hip, Knee, Trauma, Upper limb, Foot & Ankle, Paediatrics, Oncology, Spine, Arthroplasty, General