Journal of Bone and Joint Surgery - British Volume, Vol 84-B, Issue 3,
447-451.
doi: 10.1302/0301-620X.84B3.11344 Copyright © 2002 by British Editorial Society of Bone and Joint Surgery Comparison of polyethylene wear with femoral heads of 22 mm and 32 mmA PROSPECTIVE, RANDOMISED STUDYS. Eggli, MD1; S. zBrun, MD1; C. Gerber, MD2; and R. Ganz, MD1
1 Department of Orthopaedic Surgery, University of Berne, Inselspital, CH-3010 Berne, Switzerland. Correspondence should be sent to Dr S. Eggli. In this prospective, randomised study, we have compared the wear rate of cemented, acetabular polyethylene cups articulating with either a 22 mm or a 32 mm cobalt-chromium head. We evaluated 89 patients who had a total of 484 radiographs. The mean follow-up period was 71.4 months (SD 29.1). All the radiographs were digitised and electronically measured. The linear wear rate was significantly higher during the first two years and decreased after this period to a constant value. We suggest that this is partly due to a run-in process caused by irregularities between surfaces of the cup and head and an initial plastic deformation of the polyethylene. The mean volumetric wear was 120.3 mm3/year for the 32 mm head, which was significantly higher than the 41.5 mm3/year for the 22 mm heads. The mean linear wear rate was not significantly different. We were, however, unable to find radiological signs of osteolysis in the patients who had higher wear rates.
|
|