Logo of The Journal of Bone & Joint Surgery (Br)
Quick search:        
          Advanced Search
Guest Access | Sign In
Journal of Bone and Joint Surgery - British Volume, Vol 85-B, Issue 4, 504-509.
doi: 10.1302/0301-620X.85B4.13397  
Copyright © 2003 by British Editorial Society of Bone and Joint Surgery
This Article
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow My Folders
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Hernigou, P.
Right arrow Articles by Bahrami, T.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Hernigou, P.
Right arrow Articles by Bahrami, T.

Zirconia and alumina ceramics in comparison with stainless-steel heads

POLYETHYLENE WEAR AFTER A MINIMUM TEN-YEAR FOLLOW-UP

P. Hernigou, MD, Professor of Orthopaedic Surgery; and T. Bahrami, MD, Resident

University Paris XII, Hôpital Henri Mondor, 94010 Creteil, France.

Correpondence should be sent to Professor P. Hernigou.

Although alumina has been used in orthopaedic surgery since the 1970s, the long-term clinical results of zirconia have not been well documented in vivo. We studied hips with these two different ceramics during the same period and with a minimum follow-up of ten years. Because the size of the alumina and zirconia heads was different, hips with 32 mm alumina heads and those with 28 mm zirconia heads were compared with control hips with stainless-steel heads of the same size. Our aim was to compare the two ceramics.

There was an increased linear rate of penetration of the femoral heads into the liner between years five and 12 for the zirconia and the stainless-steel groups. This was severe in the zirconia group (0.4 mm/year compared with 0.13 mm/year for the stainless-steel group). During the same 12-year period there was, however, no significant change in the rate of wear in the alumina group (0.07 mm/year). The mean wear at the most recent follow-up was 1360 mm3 for the 28 mm zirconia group, 683 mm3 for the 28 mm stainless-steel group, 755 mm3 for the 32 mm alumina group and 1314 mm3 for the 32 mm stainless-steel group. The monoclinic content rose on the surface of three zirconia heads which were retrieved at revision. This change was associated with an increase in the surface roughness. A change in the roundness with an increase in the sphericity deviation was also observed both in the articular and non-articular parts of the femoral heads. The increase in rate of wear in the zirconia group was only evident after eight years and may be linked to a long-term biodegradation of zirconia in vivo, associated with the altered roughness and roundness which was observed on the retrieved heads.




This article has been cited by other articles:


Home page
J Bone Joint Surg BrHome page
Y.-H. Kim and J.-S. Kim
Tribological and material analyses of retrieved alumina and zirconia ceramic heads correlated with polyethylene wear after total hip replacement
J Bone Joint Surg Br, June 1, 2008; 90-B(6): 731 - 737.
[Abstract] [Full Text] [PDF]


Home page
J Bone Joint Surg BrHome page
T. von Schewelov, L. Sanzen, I. Onsten, A. Carlsson, and J. Besjakov
Total hip replacement with a zirconium oxide ceramic femoral head: A RANDOMISED ROENTGEN STEREOPHOTOGRAMMETRIC STUDY
J Bone Joint Surg Br, December 1, 2005; 87-B(12): 1631 - 1635.
[Abstract] [Full Text] [PDF]



(c) British Editorial Society of Bone and Joint Surgery All Rights Reserved
Registered charity no: 209299     Print ISSN: 0301-620X
Hip, Knee, Trauma, Upper limb, Foot & Ankle, Paediatrics, Oncology, Spine, Arthroplasty, General