Journal of Bone and Joint Surgery - British Volume, Vol 85-B, Issue 8,
1114-1119.
doi: 10.1302/0301-620X.85B8.13590 Copyright © 2003 by British Editorial Society of Bone and Joint Surgery Acetabular revision in total hip replacement with a press-fit cupR. Elke, MD, Orthopaedic Surgeon; B. Berli, MD, Research Fellow; A. Wagner, MD, Research Fellow; and E. W. Morscher, MD, Orthopaedic SurgeonThe Orthopaedic Department, University of Basel, CH-4031 Basel, Switzerland. Correspondence should be sent to Dr R. Elke at the Orthopaedic Department, Kantonsspital Olten, Baslesstrasse 150, CH-4600 Olten, Switzerland. The need for supplementary screw fixation in acetabular revisions is still widely debated. We carried out 439 acetabular revisions over an eight-year period. In 171 hips with contained or small segmental defects, the Morscher press-fit cup was used. These revisions were followed prospectively. No screws were used for additional fixation. A total of 123 hips with a mean follow-up of 7.4 years (5 to 10.5) were available for clinical and radiological review. There was no further revision of a press-fit cup for aseptic loosening. Radiological assessment revealed osteolysis in three hips. Of the original 171 hips there was cranial and medial migration of up to 6 mm at two years in 44 (26%). No further migration was seen after the second post-operative year. Acetabular revision without screws is possible with excellent medium-term results in well selected patients.
|
|