Journal of Bone and Joint Surgery - British Volume, Vol 86-B, Issue 5,
643-647.
doi: 10.1302/0301-620X.86B5.14524 Copyright © 2004 by British Editorial Society of Bone and Joint Surgery Total hip arthroplasty using the Wroblewski golf ball cup inserted through the posterior approachA HIGH RATE OF DISLOCATIONP. Porter, FRCS (Trauma and Orth), Specialist Registrar in Orthopaedics; and M. H. Stone, FRCS, Consultant Orthopaedic SurgeonDepartment of Orthopaedics, Leeds General Infirmary, Great George Street, Leeds, LS1 3EX, UK. Correspondence should be sent to Mr P. Porter at 6 Ullswater Drive, Wetherby, West Yorkshire LS22 6YF, UK. The Wroblewski golf ball acetabular cup was introduced by surgeons using the trochanteric osteotomy approach for revision total hip replacement (THR) in order to reduce the rate of dislocation. We have routinely used the Ogee long posterior wall (Ogee LPW) and the Wroblewski angle bore cups in THR. Although the new Wroblewski golf ball cup performed well there was a significant early rate of dislocation of 20%. Our rate of dislocation over a period of ten years using the Ogee LPW and Wroblewski angle bore cups had been 0.52%. We present our findings and an investigation as to why the new cup has such a high rate of dislocation when used with the posterior approach. We show that a relatively small change in the design of the acetabular component resulted in significant adverse clinical results. This article has been cited by other articles: