Logo of The Journal of Bone & Joint Surgery (Br)
Joint Replacement Instrumentation Limited (JRI) Ad
Quick search:        
          Advanced Search
Guest Access | Sign In
Journal of Bone and Joint Surgery - British Volume, Vol 86-B, Issue 5, 643-647.
doi: 10.1302/0301-620X.86B5.14524  
Copyright © 2004 by British Editorial Society of Bone and Joint Surgery
This Article
Right arrow Full Text (PDF)
Right arrow FrenchSpanishItalianRomanianPolishRussianCzech
Right arrow Submit a response
Right arrow View responses
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow My Folders
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Porter, P.
Right arrow Articles by Stone, M. H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Porter, P.
Right arrow Articles by Stone, M. H.

Total hip arthroplasty using the Wroblewski golf ball cup inserted through the posterior approach

A HIGH RATE OF DISLOCATION

P. Porter, FRCS (Trauma and Orth), Specialist Registrar in Orthopaedics; and M. H. Stone, FRCS, Consultant Orthopaedic Surgeon

Department of Orthopaedics, Leeds General Infirmary, Great George Street, Leeds, LS1 3EX, UK.

Correspondence should be sent to Mr P. Porter at 6 Ullswater Drive, Wetherby, West Yorkshire LS22 6YF, UK.

The Wroblewski golf ball acetabular cup was introduced by surgeons using the trochanteric osteotomy approach for revision total hip replacement (THR) in order to reduce the rate of dislocation. We have routinely used the Ogee long posterior wall (Ogee LPW) and the Wroblewski angle bore cups in THR. Although the new Wroblewski golf ball cup performed well there was a significant early rate of dislocation of 20%. Our rate of dislocation over a period of ten years using the Ogee LPW and Wroblewski angle bore cups had been 0.52%. We present our findings and an investigation as to why the new cup has such a high rate of dislocation when used with the posterior approach. We show that a relatively small change in the design of the acetabular component resulted in significant adverse clinical results.




This article has been cited by other articles:


Home page
J Bone Joint Surg BrHome page
R. J. K. Khan, D. Fick, R. Alakeson, S. Haebich, M. de Cruz, B. Nivbrant, and D. Wood
A constrained acetabular component for recurrent dislocation
J Bone Joint Surg Br, July 1, 2006; 88-B(7): 870 - 876.
[Abstract] [Full Text] [PDF]

eLetters:

Read all eLetters

Letter from Professor Wroblewski
B M Wroblewski
J Bone Joint Surg Br Online, 2 Dec 2004 [Full text]
Response to Professor Wroblewski from the authors