Logo of The Journal of Bone & Joint Surgery (Br)
Joint Replacement Instrumentation Limited (JRI) Ad
Quick search:        
          Advanced Search
Guest Access | Sign In
Journal of Bone and Joint Surgery - British Volume, Vol 90-B, Issue 5, 556-561.
doi: 10.1302/0301-620X.90B5.20545  
Copyright © 2008 by British Editorial Society of Bone and Joint Surgery
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow My Folders
Right arrow Download to citation manager
Google Scholar
Right arrow Articles by Glyn-Jones, S.
Right arrow Articles by Murray, D. W.
PubMed
Right arrow PubMed Citation
Right arrow Articles by Glyn-Jones, S.
Right arrow Articles by Murray, D. W.

The creep and wear of highly cross-linked polyethylene

A THREE-YEAR RANDOMISED, CONTROLLED TRIAL USING RADIOSTEREOMETRIC ANALYSIS

S. Glyn-Jones, FRCS (Orth), Clinical Research Fellow1; P. McLardy-Smith, FRCS, Consultant Orthopaedic Surgeon1; H. S. Gill, DPhil, Lecturer in Orthopaedic Engineering1; and D. W. Murray, FRCS(Orth), Professor of Orthopaedic Surgery1

1 Nuffield Department of Orthopaedic Surgery, University of Oxford, Nuffield Orthopaedic Centre, Windmill Road, Oxford OX3 7LD, UK.

Correspondence should be sent to Dr H. S. Gill; e-mail: richie.gill{at}ndos.ox.ac.uk

The creep and wear behaviour of highly cross-linked polyethylene and standard polyethylene liners were examined in a prospective, double-blind randomised, controlled trial using radiostereometric analysis.

We randomised 54 patients to receive hip replacements with either highly cross-linked polyethylene or standard liners and determined the three-dimensional penetration of the liners over three years.

After three years the mean total penetration was 0.35 mm (SD 0.14) for the highly cross-linked polyethylene group and 0.45 mm (SD 0.19) for the standard group. The difference was statistically significant (p = 0.0184). From the pattern of penetration it was possible to discriminate creep from wear. Most (95%) of the creep occurred within six months of implantation and nearly all within the first year. There was no difference in the mean degree of creep between the two types of polyethylene (highly cross-linked polyethylene 0.26 mm, SD 0.17; standard 0.27 mm, SD 0.2; p = 0.83). There was, however, a significant difference (p = 0.012) in the mean wear rate (highly cross-linked polyethylene 0.03 mm/yr, SD 0.06; standard 0.07 mm/yr, SD 0.05). Creep and wear occurred in significantly different directions (p = 0.01); creep was predominantly proximal whereas wear was anterior, proximal and medial.

We conclude that penetration in the first six months is creep-dominated, but after one year virtually all penetration is due to wear. Highly cross-linked polyethylene has a 60% lower rate of wear than standard polyethylene and therefore will probably perform better in the long term.






(c) British Editorial Society of Bone and Joint Surgery All Rights Reserved