Electronic Letters to:
-
- Hip:
B. Derbyshire and M. L. Porter
- A study of the Elite Plus femoral component using radiostereometric analysis
J Bone Joint Surg Br 2007; 89-B: 730-735
[Abstract]
[Full text]
[PDF]
|
|
Electronic letters published:
-
Authors' Reply
- Brian Derbyshire, Martyn L. Porter
(6 September 2007)
-
A study of the Elite Plus femoral component using radiostereometric analysis
- Michael R Whitehouse
(23 August 2007)
|
Authors' Reply |
6 September 2007 |
|
|
Brian Derbyshire, Research Fellow Wrightington Hospital, Martyn L. Porter
Send letter to journal:
Re: Authors' Reply
Brian.Derbyshire{at}wwl.nhs.uk Brian Derbyshire, et al.
|
Sir,
We thank Mr Whitehouse for his comments on our paper. He raised a
number of points which he suggested might render our conclusions
unreliable.
He questioned if we had used a calibration cage and whether this had
been validated. He also asked what size of markers we had used because he
believed this might have influenced the accuracy of our measurements. As
mentioned in our paper, we used the Swedish UmRSA system. This is
probably the most well known of RSA systems and numerous papers have been
published on its accuracy and clinical use. The system comprises
software, a validated calibration cage, a marker bead injector and
tantalum marker beads (available in diameters of 0.5 mm, 0.8 mm and 1.0
mm). The software included automatic detection of each marker bead image
centre. The two marker beads fixed to each femoral component had
diameters of 0.8 mm. We used 1.0 mm diameter beads in the proximal femur.
This size is amongst the three "standard" bead sizes described by Valstar
et al1 in reference 2 of Mr Whitehouse's letter. Indeed, 1.0 mm diameter beads
are used by Valstar’s group.2
Although we had decided upon a sample size of 25, we eventually
inserted a total of 29 RSA femoral stems (29 patients). This was because
the radiographs of four of our early patients had an insufficient number
of visible marker beads or the marker bead scatter was insufficient. A
minimum of three bone marker beads must be visible in order to determine
the 3-D position of the bone segment. If the bone markers are too close
to one another (low scatter), the accuracy of the measurement –
particularly rotations – is reduced. The UmRSA software displays a
measure of the marker bead scatter, referred to as the Condition Number.
In general, the accuracy deteriorates with increasing Condition Number.
The paper cited by Mr Whitehouse1 describes the recommended limit for the
Condition Number. The four patients were rejected during the first few
months of the project and had no bearing on the project or on our
conclusions.
We agree that double examinations should ideally have been performed
on all patients. However, there were ethical considerations relating to
the total X-ray exposure over the three year course of the project. This
was particularly true for patients who had one or more repeated X-rays
(due to inadequate radiographs) at one or more of the examination
appointments. Although the overall X-ray exposure for an RSA examination
is much less than for a standard A-P pelvic X-ray examination, the total
exposure, including repeats, over a three year period (7 appointments) can
add up, and we took due consideration of this. RSA requires a special
radiographic set up of two X-ray tubes and a calibration table. The
height of the X-ray table is adjusted to the intersection of the X-ray
beams at the proximal femur. The two stem marker bead images and the
femoral head image must be visible on both radiographs. In addition, most
of the calibration table marker bead images must be visible. Not
surprisingly, the radiographer can sometimes get it wrong, especially
during the early, learning stage. The stem tip marker image might be
missing on one radiograph, or the patient’s leg might not be rotated
sufficiently to show the marker bead image at the stem shoulder. Some of
the calibration table marker images might be too over-exposed (blacked
out), or they might be obscured by the implant. The need for repeat
exposures was therefore mainly due to error in the radiographic set-up and
not due to a specific property of the patient (e.g. size or obesity). We
do not agree, therefore, that our choice of patients partaking in the
double examinations had any bearing on the precision of the measurements.
As we pointed out in our paper, our double examination results compare
favourably with those from other studies in which the UmRSA system has
been employed. The number of patients subjected to double examinations
(5) in our study compares with the number used (7) in an RSA study of the
Elite Plus by the Oxford group.3
There was another consideration with regard to double examinations.
We did not feel that it would be fair to the patient to carry out double
examinations before six months while healing was still taking place. In
carrying out a double examination, we rearranged the X-ray equipment and
asked the patient to leave the room before re-setting the equipment and
taking the second set of radiographs. It was not necessary to put the
patient through this during one of the first few RSA appointments. We
therefore carried out double examinations during the middle stages of the
project – at which time we had sufficient records of the cumulative X-ray
exposure for each patient. Of course, patients might have accrued more
exposure at the subsequent examination appointment, but the numbers were
sufficient for us to make a judgement about this.
The main conclusion of our study was that only one femoral component
was migrating excessively at three years. We were able to attribute this
excessive migration to high patient demand. Even if our defence of the
accuracy of the RSA technique is disregarded, it is clear that there was
just one outlier in our results. That stem was continuing to migrate at
three years and the magnitude of its migration was several times that of
the other stems. The most inaccurate of techniques would have identified this.
B. DERBYSHIRE, BSc, MSc, PhD, Research Fellow,
M.L. PORTER, FRCSEd, FRCSEd(Orth), Consultant Orthopaedic Surgeon,
Wrightington Hospital,
Appley Bridge,
Wigan, UK.
1. Valstar ER, Gill R, Ryd L, et al. Guidelines
for standardization of radiostereometry (RSA) of implants. Acta Orthop 2005;76:563-72.
2. Nelissen RG, Doets HC, Valstar ER. Early migration of the tibial
component of the buechel-pappas total ankle prosthesis. Clin Orthop Relat Res 2006;448:146-51.
3. Alfaro-Adrian J, Gill H, Murray DW. Should total hip arthroplasty
femoral components be designed to subside? A radiostereomentric analysis of the Charnley Elite and Exeter stems. J Arthroplasty 2001;16:598-606. |
|
A study of the Elite Plus femoral component using radiostereometric analysis |
23 August 2007 |
|
|
Michael R Whitehouse, Specialist Registrar in Orthopaedics Royal Surrey County Hospital, Guildford, GU2 7XX
Send letter to journal:
Re: A study of the Elite Plus femoral component using radiostereometric analysis
mike_whitehouse{at}yahoo.com Michael R Whitehouse
|
Sir,
I read with interest the article by Derbyshire and Porter1 in the June 2007 issue entitled "A study of the Elite Plus femoral component using radiostereometric analysis". I would like to raise a few points regarding
the use of the radiostereometric analysis (RSA) technique in their
assessment of the motion of the Elite Plus femoral component.
There are a number of points in the paper which could make the
conclusions drawn not entirely reliable. RSA has been recognised as a
surrogate marker for final outcome with relatively small numbers of
patients over a relatively short period of follow-up.2 There have been
recent attempts to standardise the techniques used in RSA and the
presentation of results in order to allow meaningful comparisons of
studies.3 In terms of the methods used, the authors have not stated
a number of important variables. The size of the markers used has not been
expressed which can influence the accuracy of the results.3 In the case
of the four patients excluded from the study due to insufficient scatter or
number of markers visible, these variables are not stated. The authors
also do not state if a calibration cage was used and whether this was
validated.
My final concern regards the authors' use of "double examinations" in
the study. It is stated that the double examinations were only performed
on five subjects. Ideally all patients should have double examinations
performed.3 It is also stated that the patients with the lowest number
of repeat examinations had double examinations performed in order to
reduce the risk of exposure to radiation. This raises two points; it is
hard to see how the patients with the lowest number of repeat exposures
could be identified in a prospective study when the number of exposures
that would be required by subsequent subjects is not known. It is also
likely that the patients that required the least number of exposures in
order to obtain adequate radiographs are those least likely to suffer from
inaccuracies during repeat examination. This data could therefore prove
falsely reassuring.
The authors state that their results differ from those of a previous
RSA study4 and that there have been a number of conflicting clinical
studies.5-11 The data presented does add to what is already known
regarding the Elite Plus stem but it is possible that small errors or
inconsistencies introduced by the use of the RSA technique could
be important in determining the long-term results of this stem when using
a surrogate marker such as RSA.
M.R. WHITEHOUSE, MRCS(Eng) BSc,
Specialist Registrar in Orthopaedics,
Royal Surrey County Hospital,
Guildford, UK.
1. Derbyshire B, Porter ML. A study of the Elite Plus femoral component using radiostereometric analysis. J Bone Joint Surg [Br] 2007;89-B:730-5.
2. Kärrholm J, Borssén B, Löwenhielm G, Snorrason F. Does early
micromotion of femoral stem prostheses matter? 4-7-year stereoradiographic
follow-up of 84 cemented prostheses. J Bone Joint Surg [Br] 1994;76-B:912-7.
3. Valstar ER, Gill R, Ryd L, et al. Guidelines
for standardization of radiostereometry (RSA) of implants. Acta
Orthop 2005;76:563–72.
4. Alfaro-Adrian J, Gill HS, Murray DW. Should total hip arthroplasty
femoral components be designed to subside? A radiostereometric analysis study of the Charnley Elite and Exeter stems. J Arthroplasty 2001;16:598-606.
5. Livingston SJ, Ivory JP. Five year survivorship study of the Charnley
Elite Plus total hip replacement. J Bone Joint Surg [Br] 2003;85-B(Suppl 1):59-60.
6. Walton NP, Darrah C, Shepstone L, Donell ST, Phillips H. The Elite Plus total hip arthroplasty: the need for radiological surveillance. J Bone Joint Surg
[Br] 2005;87-B:458-62.
7. Rowsell M, Der Tavitian J, Birtwistle S, Power R. Survivorship of the
Charnley Elite Plus cemented femoral stem. Int Orthop 2005;29:214-18.
8. Hauptfleisch J, Glyn-Jones S, Beard DJ, Gill HS, Murray DW. The
premature failure of the Charnley Elite Plus stem: a confirmation of RSA predictions. J Bone Joint Surg [Br]
2006;88-B:179-83.
9. Kalairajah Y, Azurza K, Molloy S, et al. Is the Charnley evolution
working? a five-year outcome study. Acta Orthop Belg 2004;70:315-21.
10. Brokelman RB, van Loon CJ, Rijnberg WJ. Patient versus surgeon
satisfaction after total hip arthroplasty. J Bone Joint Surg [Br] 2003;85-B:495-8.
11. Kim YH, Kim JS, Yoon SH. Long-term survivorship of the Charnley Elite
Plus femoral component in young patients. J Bone Joint Surg [Br] 2007;89-B:449-54. |
|
|